Abstract:
The main of study to estimate the radiation dose for the patients chest
(CT) examination undergoing (CT) scanning .
This study in Khartoum – State Modern Medical Center the period
December2019. The data include 50 patients male and female.
Use machine CT GE OPTIMA 60slices.
Method include to measure the dose radiation which controlmany
parameters (Kvp, MAs,CTDLvol, DLP)this parameter was calculated using of
machine after complication exam done.
Use estimated (DLP) and effective does (E).
Analysis of data by (spss) In additional should result analysis the mean
effective does for achest (11.29) (mSv) the mean (DLp) 774.76 mGY/cm. and
the mean (CTDLvol) (4.620mGY) .
The Maximum Dose(29mSv) and the Minimum( 2mSv).
(CT) accounts for the majority of radiation exposure related to medical
imaging medical professionals should have a working knowledge of the
benefits and risk of medical radiation and an understand of strategies for
reducing (CT) radiation