Please use this identifier to cite or link to this item: https://repository.sustech.edu/handle/123456789/10284
Title: Evaluation of Quality Control of MRI Machines.
Other Titles: تقويم ضبط الجودة لاجهزة الرنين المغنطيسي
Authors: Mustafa,Abdelkarim Eltayeb Mohamed
Supervised -Hussein Ahmed Hasan
Keywords: College of Medical
Radiologic Science
Evaluation of Quality
Control of MRI Machines
Issue Date: 1-Oct-2011
Publisher: Sudan University of Science and Technology
Citation: Mustafa,Abdelkarim Eltayeb Mohamed.Evaluation of Quality Control of MRI Machines:Abdelkarim Eltayeb Mohamed Mustafa;Hussein Ahmed Hasan.-khartoum:Sudan University of Science and Technolog,College of Medical Radiologic Science ,2005.-67p. :ill. ;28cm.- M.Sc.
Abstract: Evaluation Of Quality control of MRI Machines is a very important.The aim of this study was to evaluate status of MRI machines by assessment the spatial resolution , SNR ,spatial linearity . The study was conducted at NFPMS Omdurman Sudan Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Centerfrom the period of January to August 2010. Results shows SNR was found to be 0.24, 0.29,0.19,0.16,0.13 for the NFPMS Omdurman Sudan ,Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively.that means Altaheel Hospital had scored the highest SNR the mater that affects the image quality positively. Uniformity was found to be 71%,79%,88%,90% and 86% atthe NFPMS Omdurman Sudan ,Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively, means Modern Medical Centrehad scored the highest uniformaty value, the matter that affects the image quality positivly. Resolution was evaluated and the result was found to be 1.1,0.5,0.35 and 0.4 at Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively,that means Modern Medical Centrehad scored the highestresolution images, the matter that affects the image quality positivly. Linearity was found to be 74.3%, 83.54%, 90.42% and 88.45% forAl-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively, that means Modern Medical Centrehad scored the highest linearity measurement the matter that affects the image quality positivly
Description: Thesis
URI: http://repository.sustech.edu/handle/123456789/10284
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Evaluation of Quality Control ... .pdf
  Restricted Access
Title49.71 kBAdobe PDFView/Open Request a copy
Abstract.pdf
  Restricted Access
Abstract680.9 kBAdobe PDFView/Open Request a copy
Research.pdf
  Restricted Access
Research422.76 kBAdobe PDFView/Open Request a copy


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.