Title: Assessment of kVp accuracy for X-ray department in Khartoum State Hospitals
Other Titles: تقييم دقة الكيلوفولت لأقسام الأشعة السينية بمستشفيات ولاية الخرطوم
Authors: Babekir, Omer Mohammed Elhadi
Supervisor, -Hussein Ahmed Hassan
Keywords: Medical Radiologic Sciences
Medical Physics
Assessment of kVp accuracy
X-ray department
Issue Date: 2-Nov-2017
Publisher: Sudan University of Science and Technology
Citation: Babekir, Omer Mohammed Elhadi . Assessment of kVp accuracy for X-ray department in Khartoum State Hospitals \ Omer Mohammed Elhadi Babekir ; Hussein Ahmed Hassan .- Khartoum:Sudan University of Science & Technology,Medical Radiologic Sciences,2017.- 56 p.:ill.;28cm.-M.Sc
Abstract: This study was aimed to assessment of kVp accuracy for x-ray departments in Khartoum state. A total of 15 x-ray units in radiology departments in Khartoum state were tested by KV meter tool. In this study the final results were analyzed by Microsoft Office Excel Program, the results showed that the percentage of acceptable for x-ray machines (80%) greater than unacceptable for x-ray machines (20%). The final assessment of KVp accuracy acceptable for x-ray departments is most in Khartoum state. It is clear that implementation of QC program concerning kV accuracy is very important to the production of image quality
Description: Thesis
URI: http://repository.sustech.edu/handle/123456789/19313
Appears in Collections:Masters Dissertations : Medical Radiologic Science