Please use this identifier to cite or link to this item:
https://repository.sustech.edu/handle/123456789/19313
Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Babekir, Omer Mohammed Elhadi | - |
dc.contributor.author | Supervisor, -Hussein Ahmed Hassan | - |
dc.date.accessioned | 2017-12-10T08:54:20Z | - |
dc.date.available | 2017-12-10T08:54:20Z | - |
dc.date.issued | 2017-11-02 | - |
dc.identifier.citation | Babekir, Omer Mohammed Elhadi . Assessment of kVp accuracy for X-ray department in Khartoum State Hospitals \ Omer Mohammed Elhadi Babekir ; Hussein Ahmed Hassan .- Khartoum:Sudan University of Science & Technology,Medical Radiologic Sciences,2017.- 56 p.:ill.;28cm.-M.Sc | en_US |
dc.identifier.uri | http://repository.sustech.edu/handle/123456789/19313 | - |
dc.description | Thesis | en_US |
dc.description.abstract | This study was aimed to assessment of kVp accuracy for x-ray departments in Khartoum state. A total of 15 x-ray units in radiology departments in Khartoum state were tested by KV meter tool. In this study the final results were analyzed by Microsoft Office Excel Program, the results showed that the percentage of acceptable for x-ray machines (80%) greater than unacceptable for x-ray machines (20%). The final assessment of KVp accuracy acceptable for x-ray departments is most in Khartoum state. It is clear that implementation of QC program concerning kV accuracy is very important to the production of image quality | en_US |
dc.description.sponsorship | Sudan University of Science and Technology | en_US |
dc.language.iso | en | en_US |
dc.publisher | Sudan University of Science and Technology | en_US |
dc.subject | Medical Radiologic Sciences | en_US |
dc.subject | Medical Physics | en_US |
dc.subject | Assessment of kVp accuracy | en_US |
dc.subject | X-ray department | en_US |
dc.title | Assessment of kVp accuracy for X-ray department in Khartoum State Hospitals | en_US |
dc.title.alternative | تقييم دقة الكيلوفولت لأقسام الأشعة السينية بمستشفيات ولاية الخرطوم | en_US |
dc.type | Thesis | en_US |
Appears in Collections: | Masters Dissertations : Medical Radiologic Science |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Assessment of kVp accuracy ... .pdf | Title | 54.79 kB | Adobe PDF | View/Open |
Abstract .pdf | Abstract | 210.51 kB | Adobe PDF | View/Open |
Research .pdf | Research | 1.45 MB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.