dc.identifier.citationBabekir, Omer Mohammed Elhadi . Assessment of kVp accuracy for X-ray department in Khartoum State Hospitals \ Omer Mohammed Elhadi Babekir ; Hussein Ahmed Hassan .- Khartoum:Sudan University of Science & Technology,Medical Radiologic Sciences,2017.- 56 p.:ill.;28cm.-M.Scen_US
dc.identifier.urihttp://repository.sustech.edu/handle/123456789/19313-
dc.descriptionThesisen_US
dc.description.abstractThis study was aimed to assessment of kVp accuracy for x-ray departments in Khartoum state. A total of 15 x-ray units in radiology departments in Khartoum state were tested by KV meter tool. In this study the final results were analyzed by Microsoft Office Excel Program, the results showed that the percentage of acceptable for x-ray machines (80%) greater than unacceptable for x-ray machines (20%). The final assessment of KVp accuracy acceptable for x-ray departments is most in Khartoum state. It is clear that implementation of QC program concerning kV accuracy is very important to the production of image qualityen_US
dc.description.sponsorshipSudan University of Science and Technologyen_US
dc.language.isoenen_US
dc.publisherSudan University of Science and Technologyen_US
dc.subjectMedical Radiologic Sciencesen_US
dc.subjectMedical Physicsen_US
dc.subjectAssessment of kVp accuracyen_US
dc.subjectX-ray departmenten_US
dc.titleAssessment of kVp accuracy for X-ray department in Khartoum State Hospitalsen_US
dc.title.alternativeتقييم دقة الكيلوفولت لأقسام الأشعة السينية بمستشفيات ولاية الخرطومen_US
dc.typeThesisen_US
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Assessment of kVp accuracy ... .pdfTitle54.79 kBAdobe PDFView/Open
Abstract .pdfAbstract210.51 kBAdobe PDFView/Open
Research .pdfResearch1.45 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.