Please use this identifier to cite or link to this item: https://repository.sustech.edu/handle/123456789/11099
Title: Assessment of Exposure Factors for Conventional X-Rays Machines
Other Titles: تقدير عوامل التعرض لأجهزة الأشــعة الســينية
Authors: Abugrain, Mohamed Ahmed Abdelaziz
Supervisor-Husain Ahmed Hassan
Keywords: Radiologic Science
X-Rays Machines
Assessment of Exposure Factors for Conventional
X-Rays Machines
Issue Date: 5-Jan-2015
Publisher: Sudan University of Science and Technology
Citation: Abugrain, Mohamed Ahmed Abdelaziz . Assessment of Exposure Factors for Conventional X-Rays Machines/ Mohamed Ahmed Abdelaziz Abugrain; Husain Ahmed Hassan .- khartoum: Sudan University of Science and Technology, College of Medical Radiologic Science, 2015.- 43p. : ill.; 28cm.- M. Sc.
Abstract: This study aimed to determine the Accuracy of Exposure Factors in Conventional X-rays Machines by using multifunction meter (Kv and mAs) the main result show that there was small variation between the stated (kv and mAs) and (kv and mAs) measured by the kv meter. after the final result was analysed by (SPSS) , the result showed that 50% of x-rays machines in the acceptable range and 50% of x-rays machines in the unacceptable range, when comparison between applied and measured kilovoltage. and 90% of x-rays machines in the acceptable range and 10% of x-rays machines in the unacceptable range when comparison between sitting and measured mAs.
Description: Thesis
URI: http://repository.sustech.edu/handle/123456789/11099
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Assessment of Exposure...pdf title 205.8 kBAdobe PDFView/Open
Abstract .pdfAbstract 459.93 kBAdobe PDFView/Open
Research.pdfResearch948.02 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.