SUST Repository

Evaluation of Image Quality in Fluoroscopy Machines in Khartoum State

Show simple item record

dc.contributor.author Eltayeb, Jamal Eldeen Eltayed Mohammed
dc.contributor.author Supervisor, -Hussein Ahmed Hassan Ahmed
dc.date.accessioned 2019-05-13T07:33:33Z
dc.date.available 2019-05-13T07:33:33Z
dc.date.issued 2018-12-02
dc.identifier.citation Eltayeb, Jamal Eldeen Eltayed Mohammed . Evaluation of Image Quality in Fluoroscopy Machines in Khartoum State \ Jamal Eldeen Eltayed Mohammed Eltayeb ; Hussein Ahmed Hassan Ahmed .- Khartoum : Sudan University of Science and Technology , College of Medical Radiologic Science , 2018 .-39 p. :ill;28cm.- M.Sc en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/22604
dc.description Thesis en_US
dc.description.abstract Fluoroscopy refers to the use of an X ray beam and a suitable image receptor for viewing images of processes or instruments in the body in real time. Fluoroscopic imaging trades the high signal to noise ratio (snR) of radiography for high temporal resolution, as factors that maintain patient dose at an acceptable level must be used. The aim of this study was to perform quality control (QC) of image quality for three digital fluoroscopy units used cardiovascular and interventional a radiology procedure from Khartoum state during (July 2018 – December 2018). Measurements were based on the QC protocol developed in the frame work of European Commission (EU) DIMONDIII project. The Image quality was evaluated in terms of spatial resolution and Contrast detail delectability. Whereas image quality was assessed using HuettnerType 53spatialfrequencygrating and TOR FG18 contrast detail phantom. The results show that the limiting resolution was1.5 LP/mm for image intensifier field diameters between19 and 32 cm and this value is acceptable when compare it to European Commission (EU) DIMONDIII but the limiting resolution is less than the some similar study. The results of present study can be used as baseline for future quality assurance measurements en_US
dc.description.sponsorship Sudan University of Science & Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject Medical Radiologic Sciences en_US
dc.subject Medical Physics en_US
dc.subject Image Quality en_US
dc.subject Fluoroscopy Machines en_US
dc.title Evaluation of Image Quality in Fluoroscopy Machines in Khartoum State en_US
dc.title.alternative تقييم جودة الصورة في أجهزة الأشعة المرئية في ولاية الخرطوم en_US
dc.type Thesis en_US