dc.identifier.citationHussain, Ahmed Ibrahim Elkhateim .The Calibration of Pressure Gauges Using Pressure Balance and Direct Methods \ Ahmed Ibrahim Elkhateim Hussain ; Nafisa Badreldin Mohamed Salih .- Khartoum:Sudan University of Science & Technology,College of Science,2021.- 81 p.:ill.;28cm.-M.Sc.en_US
dc.identifier.urihttps://repository.sustech.edu/handle/123456789/28197-
dc.descriptionThesisen_US
dc.description.abstractThe calibration is defines as set of operations that establish, under specified conditions, the relationship between values of quantities indicated by a measuring instrument or measuring system, or values represented by a material measure or a reference material, and the corresponding values realized by standards. The main objective of this research is to make a comparison of calibrate of pressure gauge by using two methods, the pressure balance method and the direct comparison method. The experimental work has been carried out in pressure calibration laboratory at Sudanese Standard & Metrology Organization, calibration of pressure gauge was studied first by pressure balance method as a primary standard method and then by the direct comparison method as a secondary standard of calibration pressure indicators. Four pressure gauges were calibrated by the two methods in which each one has been measured in various measurement points and then the error value was calculated. In the first method (pressure balance method) the pressure measured after it’s generated from the weights placed on the pressure balance and correspondingly the error was evaluated with comparison to standard value, in the second method the error of the pressure gauge measurement calculated after being measured from direct pressure generation of pressure pump. The first method was found to be more accurate than the second one and that’s owing to many parameters affecting the measurement are taken into account in calculating pressure value so that the error value is more precise, therefore the pressure balance method is the primary calibration method.en_US
dc.description.sponsorshipSudan University of Science & Technologyen_US
dc.language.isoenen_US
dc.publisherSudan University of Science & Technologyen_US
dc.subjectScienceen_US
dc.subjectphysicsen_US
dc.subjectCalibrationen_US
dc.subjectPressure Gaugesen_US
dc.subjectPressure Balanceen_US
dc.titleThe Calibration of Pressure Gauges Using Pressure Balance and Direct Methods A Comparative Studyen_US
dc.title.alternativeمعايرة عدادات الضغط بإستخدام طريقتي ميزان الضغط والطرق المباشرةen_US
dc.typeThesisen_US
Appears in Collections:Masters Dissertations : Science

Files in This Item:
File Description SizeFormat 
The Calibration of .... .pdfResearch1.68 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.