Please use this identifier to cite or link to this item: https://repository.sustech.edu/handle/123456789/2152
Title: Applications of Quality Assurance for Tc 99m MAG3 & DMSA in Department of Nuclear Medicine
Authors: Abdeen, Mohamed Ahmed
Supervisor-Mohammed Elfadiel
Keywords: Medicine
Nuclear
Issue Date: 1-Jan-2009
Publisher: sudan university of science and technology
Citation: Abdeen ,Mohamed Ahmed .Applications of Quality Assurance for Tc 99m MAG3 & DMSA in Department of Nuclear Medicine/Mohamed Ahmed Abdeen ; Mohammed Elfadiel.- Khartoum : sudan university of science and technology,Medical Radiologic Science,2009 .- 96p. : ill .;28cm.- M.Sc.
Abstract: This prospective study was aimed to assess inter- and intra-operator variability during routine quality control (QC) procedure for technetium-99m (99mTc-MAG3 and DMSA) instant kit formulation. A total of 10 QC analyses with thin layer chromatography (TLC) for 20 separate MAG3 reconstitutions were performed. The percentage of free and hydrolysed 99mTc as well as binding efficiency, were calculated according to standard TLC. Each QC analysis was done using silica gel (SG), silica acrylic (SA), Whatman 1 (W1) and Whatman 3 (W3) TLC strips separately at 1h, following labeling MAG3 instant kit with 99mTc-pertechnetate. To assess the radiochemical stability of 99mTc-MAG3, the same analysis was performed after kit reconstitution. Visual confirmation for QC with scintigraphy was also performed. At both time points, each repeated all the procedure twice for each of the TLC paper types to analyze the intra-operator reliability. Crombach’s Test was used for the reliability analysis. High inter-operator correlation ratios (range: 0.821-0.920) per each TLC strip were found where the highest concordance rate was 0.921 for SA. Each TLC strip showed adequate kit reconstitution with acceptable free and hydrolysed 99mTc percentages in all analyses, along with high binding efficiency values of 94.3 ± 2.9 and 92.5 ± 1.9 at respectively. Intra-observers reliability showed almost equal high concordance rates (range: 0.888-0.961) for all types of strips. In conclusion, all kinds of ITLC/TLC strips were reliable to assess stability of the radiopharmaceutical at 1 and 4 h while analysis with the SA strip had the highest concordance rate. Inter- and intra-operator QC was also reliable.
Description: Thesis
URI: http://repository.sustech.edu/handle/123456789/2152
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Research .pdf
  Restricted Access
Research319.57 kBAdobe PDFView/Open Request a copy
appendex.pdf
  Restricted Access
appendex74.52 kBAdobe PDFView/Open Request a copy
Applications of Quality... .pdfTitle29.61 kBAdobe PDFView/Open
Abstract .pdfAbstract87.09 kBAdobe PDFView/Open
Refrences.pdf
  Restricted Access
Refrences49.41 kBAdobe PDFView/Open Request a copy


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.