https://repository.sustech.edu/handle/123456789/2122| Title: | Radiation dose measurement for patients and staff during cardiac catheterization |
| Authors: | Joda, Hiba Hashim Mohammed Supervisor - Abdelmoniem Adam Sulaiman |
| Keywords: | radiolation dosimetry |
| Issue Date: | 1-Jul-2009 |
| Publisher: | sudan university of science and technology |
| Citation: | Joda,Hiba Hashim Mohammed .Radiation dose measurement for patients and staff during cardiac catheterization/Hiba Hashim Mohammed Joda;Abdelmoniem Adam Sulaiman.- Khartoum : sudan university of science and technology,Medical Radiologic Science,2009.- 54p:ill;28cm.- M.Sc. |
| Abstract: | The primary objective of this study was to determine the patient and staff dose during cardiac catheterization procedures in Ahmed Gasim Hospital, Khartoum Bahry. A survey of patient and staff exposure was performed covered 2 Cath Lab units from 2 manufacturers. The measurements involved 50 operations. The medical staff was monitored using TLD chips (LiF: Mg, Cu, P). The main operator who was closer to the patient and the X-ray tube, was monitored at six positions (forehead, neck, chest – over the lead apron, waist – under the lead apron, leg, and hand), while the exposure to the assistant was measured at two positions (chest – over the lead apron, and hand), where the technologist and the circulator were monitored at one position (chest – over the lead apron). Patient exposure was measured using the DAP meter. The main operator and the rest of the staff received 0.14, 0.01 mSv/procedure, giving rise to an estimated annual effective dose of 17.67, 6.3 mSv/y respectively. The estimated patient dose rate was found to be 125 mGy/min which considered higher than the recommended DRL for the continuous high mode fluoroscopy used in interventional radiology (100 mGy/min). The study concluded to the fact that the main operator received relatively high dose which is a direct result to the poor radiation protection in the department. |
| Description: | Thesis |
| URI: | http://repository.sustech.edu/handle/123456789/2122 |
| Appears in Collections: | Masters Dissertations : Medical Radiologic Science |