Title: Assessment of 99mTc 99Mo Dose Calibrator Performance For Nuclear Medicine Department In Al Nelein Diagnostic Center
Other Titles: تقدير اداء جهاز قياس جرعات التكنشيوم 99م والمولبيدينوم 99 لقسم الطب النووي بمركز النيلين التشخيصي
Authors: Ali, Mosaab Mergani Hissin
Supervisor-SALAH ALI FADLALLA
Keywords: Medical Physics
Assessment of 99mTc-99Mo Dose
Calibrator Performance
Nuclear Medicine
Issue Date: 1-Mar-2015
Publisher: Sudan University of Science and Technology
Citation: Ali,Mosaab Mergani Hissin. Assessment of 99mTc 99Mo Dose Calibrator Performance For Nuclear Medicine Department In Al Nelein Diagnostic Center \Mosaab Mergani Hissin Ali;SALAH ALI FADLALLA.-khartoum :Sudan University of Science and Technology ,Medical Laboratory Science ,2015.-62p . :ill ;28cm.- M.Sc.
Abstract: Nuclear medicine uses many different radioactive isotopes for radiation diagnostics studies and for therapy. The amount of radioactivity has to be determined exactly before it is applied to a patient. The dose calibrator has to measure the radioactivity of gamma and beta with different energies precisely for high quality imaging and for applying the right amount of radiation to treat diseases. This study was carried out to assess the performance of dose calibrator which is used in nuclear medicine department in Elnielin Medical Diagnostic Center (Khartoum state). The study period was form July 2014 to February 2015. Four quality control tests were carried out using standard radionuclide, 137Cs . The tests included accuracy, constancy, linearity and geometry. All results that were obtained from the study were compared with the international standard (±5%) and the results showed that the dose calibrator under the study is of good performance and there is no need for any correction tables or factors or maintenance for the time being.
Description: Thesis
URI: http://repository.sustech.edu/handle/123456789/10877
Appears in Collections:Masters Dissertations : Medical Radiologic Science