| dc.description.abstract | The aim of the study Practical and Theoretical Measurement of 60Co Teletherapy Unit Output. The variables have been collected at (Rick) using: solid phantom (PMMA), Electrometer, Wire link, Ionization chamber (Type: FC-65P), Teletherapy cobalt-60, Barometer, Thermometer and the results have been analyzed using Excel software and showed that: the temperature effects lead to increment in the output based on the equation: , and the effect of pressure was as a reduction in the output reading and the correlation based on the equation: , where x refers to temperature/pressure in C and KPa respectively and y refers to output in mGy. Also the effect of both temperature and pressure was as an increment in the output based on the equation: where x refers to the factor of temperature/pressure and y refers to output in mGy. | en_US |
| dc.description.sponsorship | sudan university of science and technology | en_US |
| dc.language.iso | en | en_US |
| dc.publisher | sudan university of science and technology | en_US |
| dc.subject | Practical | en_US |
| dc.subject | Measurement | en_US |
| dc.title | Practical and Theoretical Measurement of 60Co Teletherapy Unit Output | en_US |
| dc.type | Thesis | en_US |