dc.contributor.author |
Alawad, Nafisa Mohamed A. |
|
dc.contributor.author |
Supervisor - Mohammed Ahmed Ali Omer |
|
dc.date.accessioned |
2014-03-17T09:17:07Z |
|
dc.date.available |
2014-03-17T09:17:07Z |
|
dc.date.issued |
2013-08-01 |
|
dc.identifier.citation |
Alawad,Nafisa Mohamed A. Practical and Theoretical Measurement of 60Co Teletherapy Unit Output/Nafisa Mohamed A. Alawad;Mohammed Ahmed Ali Omer.-Khartoum : sudan university of science and technology,Medical Radiologic science,2013.- 59p. : ill . ;28cm. M.Sc. |
en_US |
dc.identifier.uri |
http://repository.sustech.edu/handle/123456789/3949 |
|
dc.description |
Thesis |
en_US |
dc.description.abstract |
The aim of the study Practical and Theoretical Measurement of 60Co Teletherapy Unit Output.
The variables have been collected at (Rick) using: solid phantom (PMMA), Electrometer, Wire link, Ionization chamber (Type: FC-65P), Teletherapy cobalt-60, Barometer, Thermometer and the results have been analyzed using Excel software and showed that: the temperature effects lead to increment in the output based on the equation: , and the effect of pressure was as a reduction in the output reading and the correlation based on the equation: , where x refers to temperature/pressure in C and KPa respectively and y refers to output in mGy. Also the effect of both temperature and pressure was as an increment in the output based on the equation: where x refers to the factor of temperature/pressure and y refers to output in mGy. |
en_US |
dc.description.sponsorship |
sudan university of science and technology |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
sudan university of science and technology |
en_US |
dc.subject |
Practical |
en_US |
dc.subject |
Measurement |
en_US |
dc.title |
Practical and Theoretical Measurement of 60Co Teletherapy Unit Output |
en_US |
dc.type |
Thesis |
en_US |