dc.contributor.author |
Eltingari, Elsir Siddig Hamid |
|
dc.contributor.author |
Supervisor -Mohamed
E
lfadi
l
Mohamed |
|
dc.date.accessioned |
2014-03-09T12:32:53Z |
|
dc.date.available |
2014-03-09T12:32:53Z |
|
dc.date.issued |
2013-08-01 |
|
dc.identifier.citation |
Eltingari,Elsir Siddig Hamid.Development of step wedge phantom for Dental X ray machine Quality Control test/Elsir Siddig Hamid Eltingari;Mohamed Elfadil Mohamed.- Khartoum : sudan university of science and technology, Medical Radiologic Science, 2013.-50p. : ill. ;28cm. M.sc. |
en_US |
dc.identifier.uri |
http://repository.sustech.edu/handle/123456789/3831 |
|
dc.description |
Thesis |
en_US |
dc.description.abstract |
ABSTRACT
The aim of this study was to development step wedge dental phantom which calibrated with the standard phantom for accuracy fitting; that used in quality control in dental x-ray radiography and film automatic processing machine. The phantom after has been checked and assimilated, were used in Quality Control for dental intra-oral machine. The result of this study showed that application of design phantom in QC showed similar Result as Standard Phantom. T-test was applied to test the significant difference between the two phantoms at P =0.05 in case of transmission using air, Cu and Pb where t-value were equal to 0.33, 0.07 and 1.6 Respective with p equal to 0.7 ,0.9and 0.2 respectively .
Optical density also showed insignificant different between the two phantoms at p =0.05 with t value = 0.02 and p =045. The optical density result also showed an exponential response of absorbed radiation versus thickness of step wedge for design phantom it was y=0.8588+e0.0064
R2=0.7861 and for the standard phantom was y=0.7801+e 0.0456, R2=0.8978. |
en_US |
dc.description.sponsorship |
sudan university of science and technology |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
sudan university of science and technology |
en_US |
dc.subject |
Development |
en_US |
dc.subject |
Dental X- |
en_US |
dc.title |
Development of step-wedge phantom for Dental X-ray machine Quality Control test |
en_US |
dc.type |
Thesis |
en_US |