Ahmed,Omima Adam Sabil .Assessment of Co-57 image uniformity tests for gamma camera system/Omima Adam Sabil Ahmed;Mohamed Ahmed Ali Omer-Khartoum : sudan university of science and technology,Medical Radiologic Science,2010-30p. : ill .;28cm .- M.Sc. |
en_US |
dc.identifier.uri |
http://repository.sustech.edu/handle/123456789/2822 |
|
dc.description |
Thesis |
en_US |
dc.description.abstract |
Extracorporeal shock wave lithotripsy is considering the extensive world
external treatment process for renal, uretric , bladder and gallbladder
stones. This procedure is usually followed up by using x-ray fluoroscopy
devices as localizer, the reason is that to optimize the absorbed dose of
patients according to the recommendations of international organizations;
the aim of this study is to measure the entrance surface dose and the
probability of carcinogenesis regarding to x-ray fluoroscopy irradiating
during ESWL. The study was applied in Al-Naileen diagnostic center
(Group 1, 50 patient) and in Al-Khartoum advance diagnostic center
(Group 2, 25 patients), the classification of patients was depending on
the type of the lithotripter. The entrance surface dose measurements were
done by using thermoluminescence dosimeter (TLDs) GR200A LiF. The
mean entrance surface dose and mean effective dose were (0.38 mGy)
and (0.34 mGy) for group1 and 2 respectively, the measured organ dose
were (0.16 mGy), (0.02) and (0.01) for kidneys, intestines and ovaries. As
it was observed Group1 was irradiated to higher dose than Group 2 that’s
because of the x-ray fluoroscopy equipment in Group 1 has no ability to
change the orientation of the machine without exposing the patient to
radiation, thus the duration of irradiating is extended and as a result then
increasing the entrance surface dose. The results show that the
probability of carcinogenesis due to irradiating to the measured amounts
of radiation exposures is a tiny value (1 for million patient) but the main
biological effect is occur due to the cumulative impact of radiation. In
addition the study insures that there is a correlation between the weights,
irradiating factors and absorbed doses. As a result guiding charts, training
courses for technologist and a strong quality assurance program were
recommended to optimize the ESD for patients. |
en_US |
dc.description.sponsorship |
sudan university of science and technology |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
sudan university of science and technology |
en_US |
dc.subject |
Measurement |
en_US |
dc.subject |
radiation |
en_US |
dc.title |
Assessment of Co-57 image uniformity tests for gamma camera system |
en_US |
dc.type |
Thesis |
en_US |