Evaluation of Thyroid Uptake of Technetium-99m Using Gamma Camera and Dose Calibrator Methods in the National Cancer Institute - Wad Madani

Show simple item record

dc.contributor.author Abdelrahman, ZobaiTageldinYousif
dc.contributor.author Supervisor, -Salah Ali Fadlallah
dc.date.accessioned 2021-04-04T09:47:14Z
dc.date.available 2021-04-04T09:47:14Z
dc.date.issued 2020-12-02
dc.identifier.citation Abdelrahman, ZobaiTageldinYousif . Evaluation of Thyroid Uptake of Technetium-99m Using Gamma Camera and Dose Calibrator Methods in the National Cancer Institute - Wad Madani \ ZobaiTageldinYousifAbdelrahman ; Salah Ali Fadlalla .- Khartoum:Sudan University of Science and Technology,College of Medical Radiologic Science,2020.- 56 p.:ill.;28cm.-M.Sc en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/25992
dc.description Thesis en_US
dc.description.abstract The main objective of this study was to evaluate the thyroid uptake of technetium-99m using gamma camera and dose calibrator methods in National Cancer Institute. This study included 64 patients (90.6% females and 9.4% male) in the age range between 13 - 81 years, with 43.8% of them in age group 31- 48 years and mean of 37.25±14.58 years. The study was done using two methods for dose calibration, gamma camera planar system with pinhole collimator and dose calibrator at distance of 5cm and 7cm from collimator to patient's neck. The duration of the study was from March 2019 to July 2019. The study showed that the thyroid uptake values using dose calibrator method were more consistent with radiation protection principles of minimizing the exposure to radiation for staff. The method was also had higher values than those measured with gamma camera due to the scatter radiation in the method of gamma camera. Finally, the study revealed that the correlation between the two methods was given by the R-Square of 0.9937 at 5cm distance and 0.9591 at 7cm distance. According to this coefficient there was no big difference in uptake between the two methods especially at 5cm object-to-pinhole distance, which gave the best result for optimum imaging in 99mTc-pertechnetate thyroid uptake test due to minimum distance. en_US
dc.description.sponsorship Sudan University of Science & Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject Medical Radiologic Sciences en_US
dc.subject Nuclear Medicine Technology en_US
dc.subject Evaluation of Thyroid Uptake en_US
dc.subject Technetium-99m en_US
dc.subject Dose Calibrator Methods in the National Cancer Institute en_US
dc.title Evaluation of Thyroid Uptake of Technetium-99m Using Gamma Camera and Dose Calibrator Methods in the National Cancer Institute - Wad Madani en_US
dc.title.alternative تقويم إمتصاص الغدة الدرقية للتكنيشيوم 99م باستخدام طريقتي القاما كاميرا وقياس الجرعات بالمعهد القومي للسرطان – ود مدني en_US
dc.type Thesis en_US