Fadlalmola, Rawda Abd.alateef . Assessment of Grid Alignment of X-Ray Units in Khartoum State \ Rawda Abd.alateef Fadlalmola ; Hussein Ahmad Hassan .- Khartoum : Sudan University of Science and Technology , College of Medical Radiologic Science , 2018.- 51p. :ill;28cm.- M.Sc |
en_US |
dc.identifier.uri |
http://repository.sustech.edu/handle/123456789/23591 |
|
dc.description |
Thesis |
en_US |
dc.description.abstract |
This study aimed to assess grid alignment for radiology departments in Khartoum state.
A total of 10 x-ray units in radiology departments in Khartoum state from different manufacture were tested by grid alignment test tool and the optical density for radiographic film was read by transition densitometer .
In this study the final result showed that the percentage of grid misalignment for x-ray machines was (40%), less than those with grid alignment x-ray machines (60%).
Which means almost of the grid were found misalignment.
The study recommended that all x-ray equipment should be regularly checked using quality assurance tests. |
en_US |
dc.description.sponsorship |
Sudan University of Science & Technology |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
Sudan University of Science and Technology |
en_US |
dc.subject |
Medical Radiologic Sciences |
en_US |
dc.subject |
Medical Physics |
en_US |
dc.subject |
Grid Alignment |
en_US |
dc.subject |
Grid Alignment of X-Ray |
en_US |
dc.title |
Assessment of Grid Alignment of X-Ray Units in Khartoum State |