| dc.identifier.citation | Fadlalmola, Rawda Abd.alateef . Assessment of Grid Alignment of X-Ray Units in Khartoum State \ Rawda Abd.alateef Fadlalmola ; Hussein Ahmad Hassan .- Khartoum : Sudan University of Science and Technology , College of Medical Radiologic Science , 2018.- 51p. :ill;28cm.- M.Sc | en_US |
| dc.identifier.uri | http://repository.sustech.edu/handle/123456789/23591 | |
| dc.description | Thesis | en_US |
| dc.description.abstract | This study aimed to assess grid alignment for radiology departments in Khartoum state. A total of 10 x-ray units in radiology departments in Khartoum state from different manufacture were tested by grid alignment test tool and the optical density for radiographic film was read by transition densitometer . In this study the final result showed that the percentage of grid misalignment for x-ray machines was (40%), less than those with grid alignment x-ray machines (60%). Which means almost of the grid were found misalignment. The study recommended that all x-ray equipment should be regularly checked using quality assurance tests. | en_US |
| dc.description.sponsorship | Sudan University of Science & Technology | en_US |
| dc.language.iso | en | en_US |
| dc.publisher | Sudan University of Science and Technology | en_US |
| dc.subject | Medical Radiologic Sciences | en_US |
| dc.subject | Medical Physics | en_US |
| dc.subject | Grid Alignment | en_US |
| dc.subject | Grid Alignment of X-Ray | en_US |
| dc.title | Assessment of Grid Alignment of X-Ray Units in Khartoum State | en_US |
| dc.type | Thesis | en_US |