SUST Repository

Assessment of Dose Calibrator Performance in the Nuclear Medicine Department of Elnilein Medical Diagnostic Centre

Show simple item record

dc.contributor.author Maki, Eshraga Abubaker
dc.contributor.author Supervisor, -Awad Abdalla Adlan
dc.date.accessioned 2018-01-02T07:12:40Z
dc.date.available 2018-01-02T07:12:40Z
dc.date.issued 2017-10-25
dc.identifier.citation Maki, Eshraga Abubaker . Assessment of Dose Calibrator Performance in the Nuclear Medicine Department of Elnilein Medical Diagnostic Centre \ Eshraga Abubaker Maki ; Awad Abdalla Adlan .- Khartoum : Sudan University of Science & Technology,Medical Radiologic Sciences,2017.- 67 p.:ill.;28cm.- M.Sc en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/19747
dc.description Thesis en_US
dc.description.abstract Nuclear medicine uses many different radioactive isotopes for radiation diagnostic studies and for therapy . The amounts of radioactivity has to be determined exactly before it is applied to patients. The dose calibrator has to measure the radioactivity of gamma and beta with different energies precisely for high quality imaging and for applying the right amount of radiation to have good images which give good diagnosis and effective treatment of diseases This study was carried out to assess the performance of dose calibrator that is working in Nuclear medicine department of Elnilein Medical Diagnostic Center in Khartoum State. Four quality control tests were carried out using two standard radionuclides Cs137 and Co57. Tests included accuracy linearity, constancy and geometry. All results that were obtained from the study were compared with the international standard (± 10%) and results showed that the dose calibrator has good performance and there is no need for any correction tables or factors or maintenance for the time being. en_US
dc.description.sponsorship Sudan University of Science & Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject Medical Radiologic Sciences en_US
dc.subject Nuclear Medicine Technology en_US
dc.subject Dose Calibrator Performance en_US
dc.subject Elnilein Medical Diagnostic Centre en_US
dc.title Assessment of Dose Calibrator Performance in the Nuclear Medicine Department of Elnilein Medical Diagnostic Centre en_US
dc.title.alternative تقويم أداء جهازقياس جرعات النشاط الإشعاعي بقسم الطب النووي بمركز النيلين للتشخيص الطبي بالخرطوم en_US