SUST Repository

Automatic Analysis’s of SPECT Gamma Camera performance using Image Processing

Show simple item record

dc.contributor.author Ahmed, Einas Mohamed
dc.contributor.author Supervisor, -Mohamed Elfadil
dc.date.accessioned 2017-11-28T09:55:07Z
dc.date.available 2017-11-28T09:55:07Z
dc.date.issued 2017-09-11
dc.identifier.citation Ahmed, Einas Mohamed . Automatic Analysis’s of SPECT Gamma Camera performance using Image Processing \ Einas Mohamed Ahmed ; Mohamed Elfadil .- Khartoum:Sudan University of Science & Technology,College of Medical Radiologic Science,2017.- 110 p.:ill.;28cm.-Ph.D en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/19162
dc.description Thesis en_US
dc.description.abstract The main objective of this study was to assess the quality control of SPECT gamma camera using developed Algorithm via image processing procedure. Usually assessment done through visual inspection of phantom images i.e. visual perception trend; in this study separate algorithm were developed to quantify the uniformity (integral, and differential), linearity and resolution using Modulation Transfer Function (MTF). The data of this study were collected from Royal Care International Hospital, Nuclear Medicine Department in the period from 2014 to 2016; images of QC test were taken from SPECT gamma camera as DICOM format. The result of this study showed that the uniformity for useful central field of view for integral uniformity was 2.97%. While row differential uniformity was 1.96% and column differential uniformity was 1.55% similar to build in routine. The linearity showed that for vertical location the variation from the original location of the line it was 0.02 mm which indicates that almost the point’s falls in the line with a correlation coefficient of 0.998. For horizontal lines the differences between the two in average equal to 0.2mm, with a correlation coefficient equal to 0.956. The resolution for minimum object used 4mm was 70% and increased with object size increased to reach 100% resolution for an object size of 10 mm i.e. 100% resolution corresponds to the coordinate of the MTF at 10% and frequency of 0.05cycle /mm. in conclusion the developed algorithm can be used to assess all gamma camera QC image regardless of the machine brand objectively en_US
dc.description.sponsorship Sudan University of Science and Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject Medical Radiologic Sciences en_US
dc.subject Medical Physics en_US
dc.subject Automatic Analysis’s en_US
dc.subject SPECT Gamma Camera performance en_US
dc.title Automatic Analysis’s of SPECT Gamma Camera performance using Image Processing en_US
dc.title.alternative التحليل الآلي لأداء جهاز الجاما كاميرا باستخدام معالجة الصور en_US
dc.type Thesis en_US