Abstract:
The main objective of this study was to assess patient dose during x-ray examination of skull and thoraco-lumbar in Red Sea state Hospitals. The dose received by 146 adult patients was measured using CALDOSE software by entering the parameters: focus to detector distance FDD, mAs , KVP.
The result of the study showed that the mean of ESAK were found to be (0.79 ± 0.30) for skull and (0.37 ± 0.2) for thoraco-lumbar. The INAK were found to be (0.52 ± 0.14) for thoracic lumbar and (0.67 ± 0.25) for skull. The RCI were found to be (0.39 ± 0.19) for skull and (1.04 ± 0.75) for thoraco-lumbar.
The mean ESAK values obtained are found to be within the standardreference. The data obtained may add to the available information in nationalrecords for general use. It may provide guidance on where efforts on dosereduction will need to be directed to fulfill the requirements of the optimizationprocess and serve as a reference for future researches.