Assessment of Grid Alignment for X-Ray Units in Khartoum State

Show simple item record

dc.contributor.author Mutwakil, Menas Abd.almonem
dc.contributor.author Supervised,- Hussein Ahmad Hassan
dc.date.accessioned 2016-06-13T07:36:34Z
dc.date.available 2016-06-13T07:36:34Z
dc.date.issued 2016-03-06
dc.identifier.citation Menas Abd.almonem.Assessment of Grid Alignment for X Ray Units in Khartoum State/Menas Abd.almonem Mutwakil; Hussein Ahmad Hassan.-Khartoum:Sudan University of Science and Technology,College Radiologic Sciences,2016.-55p:ill;28cm.-M.Sc. en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/13652
dc.description Thesis en_US
dc.description.abstract This study was aimed to assessment of grid alignment for radiology departments in Khartoum state. A total of 15 x-ray units in radiology departments in Khartoum state were tested by grid alignment test tool and the optical density for radiographic film was read by transition densitometer. In this study the final result showed that the percentage of grid misalignment for x-ray machines (66.7%) greater than the grid alignment x-ray machines (33.3%). The final assessment of grid misalignment for radiology departments is most in Khartoum state. This study recommended that all x-ray equipment’s should be regularly checked using quality assurance tests. en_US
dc.description.sponsorship Sudan University of Science and Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject Radiologic Medical Sciences en_US
dc.subject X-Ray en_US
dc.subject Assessment of Grid en_US
dc.title Assessment of Grid Alignment for X-Ray Units in Khartoum State en_US
dc.type Thesis en_US