Evaluation of dose Performance calibrator in Radiation Therapy and Nuclear Medicine in Center Khartoum (Nuclear Medicine Department)

Show simple item record

dc.contributor.author Ibrahim, Osama Ibrahim Mohamed
dc.contributor.author Supervisor, - Awadh Abdallah Adlan
dc.date.accessioned 2016-03-29T09:08:42Z
dc.date.available 2016-03-29T09:08:42Z
dc.date.issued 2016-03-03
dc.identifier.citation Ibrahim, Osama Ibrahim Mohamed.Evaluation of dose Performance calibrator in Radiation Therapy and Nuclear Medicine in Center Khartoum Nuclear Medicine Department/Osama Ibrahim Mohamed Ibrahim;Awadh Abdallah Adlan.-Khartoum:Sudan University of Science and Technology,College Radiologic Sciences,2016.-70p:ill;28cm.-M.Sc. en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/13102
dc.description Thesis en_US
dc.description.abstract This study is concerned with quality control of the dose calibrator which is located in the Radiation Therapy and Nuclear Medicine Center of Khartoum .The objective of the study is to evaluate the dose calibrator performance .Quality control tests were performed in the period between Dec/ 2015 and Jan/2016 .They included general observations to the components of the device, background measurements, precision, accuracy, continuity ,linearity and geometric dependency . Two standard radionuclide’s Cs-137 and Tc-99m were utilized for the purpose of the study and has come to the following results, all the components of the device work properly and readings for background radioactivity were in the acceptable range (0.562+0.45 ) . Concerning Precision testing the error of the reading was (0.39%), and it was within the accepted range. Concerning accuracy testing the error of the readings was within (0.27% ) , and it was within the accepted range .For constancy test the error value of readings was (2.47%) , and it was also within the accepted range and the error of readings concerning linearity testing was (3.7%) , and it was within the accepted range and finally the results of geometric dependency showed that the correction factor was within the accepted range(0.95-1.05). All the results showed that the devise has good performance and there is no need for any correction and maintenance. en_US
dc.description.sponsorship Sudan University of Science and Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject Radiologic Sciences en_US
dc.subject Nuclear Medicine Department en_US
dc.subject Radiation Therapy en_US
dc.subject Center Khartoum en_US
dc.title Evaluation of dose Performance calibrator in Radiation Therapy and Nuclear Medicine in Center Khartoum (Nuclear Medicine Department) en_US
dc.title.alternative تقويم اداء جهاز قياس الجرعه الاشعاعيه بمركز الخرطوم للعلاج بالاشعه والطب النووي(قسم الطب النووي) en_US
dc.type Thesis en_US