SUST Repository

Calibration of Gammacell 220 E Irradiator Using Fricke and Alanine Dosimeters

Show simple item record

dc.contributor.author Rushdi, Mohamed .A .H.
dc.contributor.author Supervisor - Ibrahim. A. Shaddad
dc.date.accessioned 2015-02-04T09:14:01Z
dc.date.available 2015-02-04T09:14:01Z
dc.date.issued 2005-01-01
dc.identifier.citation Rushdi ,Mohamed .A .H.Calibration of Gammacell 220 E Irradiator Using Fricke and Alanine Dosimeters/ Mohamed .A .H. Rushdi ;Ibrahim. A. Shaddad.-khartuom:Sudan University of Science and Technology,College of Medical Radiologic Science,2005.-71p.:ill .;28cm.-M.-SC. en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/10287
dc.description Thesis en_US
dc.description.abstract Using of Gammacell 220 Excel irradiators is widely spread in many countries. This type of irradiators is being used for research purposes. Gammacell 220 Excel was provided by the International Atomic Energy Agency (IAEA) to the Radiation Processing laboratory of Sudan Atomic Energy Commission (SAEC). It is a self-contained gamma irradiator and self shielded, this makes it operates safely. Dose calibration for this cell is important for samples irradiation. In this work, a dosimetry system for the GC220E of SAEC was established using Fricke dosimeter. Fricke dosimeter has a confidence 95% in the range not exceed 400 Gy. To establish routine dosimetry at high doses up to 5000 Gy, Alanine dosimeter was used. This range can demonstrate the ability of GC220E to deliver known controllable doses in reproducible manner for high doses. The irradiation specifications often include a lower and upper limit of absorbed dose or central target dose. Absorbed dose mapping was carried out by both dosimeters to determine the magnitude and locations of D.max and D.min in the irradiation chamber. The results are in good agreement with dose distribution given in the machine manual. A comparison between the tow dosimeters was done and explained. en_US
dc.description.sponsorship Sudan University for Science & Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject College of Medical en_US
dc.subject Department of Physics en_US
dc.subject Alanine Dosimeters en_US
dc.subject Calibration of Gammacell 220 E en_US
dc.title Calibration of Gammacell 220 E Irradiator Using Fricke and Alanine Dosimeters en_US
dc.type Thesis en_US