SUST Repository

Evaluation of Quality Control of MRI Machines.

Show simple item record

dc.contributor.author Mustafa,Abdelkarim Eltayeb Mohamed
dc.contributor.author Supervised -Hussein Ahmed Hasan
dc.date.accessioned 2015-02-04T08:19:26Z
dc.date.available 2015-02-04T08:19:26Z
dc.date.issued 2011-10-01
dc.identifier.citation Mustafa,Abdelkarim Eltayeb Mohamed.Evaluation of Quality Control of MRI Machines:Abdelkarim Eltayeb Mohamed Mustafa;Hussein Ahmed Hasan.-khartoum:Sudan University of Science and Technolog,College of Medical Radiologic Science ,2005.-67p. :ill. ;28cm.- M.Sc. en_US
dc.identifier.uri http://repository.sustech.edu/handle/123456789/10284
dc.description Thesis en_US
dc.description.abstract Evaluation Of Quality control of MRI Machines is a very important.The aim of this study was to evaluate status of MRI machines by assessment the spatial resolution , SNR ,spatial linearity . The study was conducted at NFPMS Omdurman Sudan Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Centerfrom the period of January to August 2010. Results shows SNR was found to be 0.24, 0.29,0.19,0.16,0.13 for the NFPMS Omdurman Sudan ,Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively.that means Altaheel Hospital had scored the highest SNR the mater that affects the image quality positively. Uniformity was found to be 71%,79%,88%,90% and 86% atthe NFPMS Omdurman Sudan ,Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively, means Modern Medical Centrehad scored the highest uniformaty value, the matter that affects the image quality positivly. Resolution was evaluated and the result was found to be 1.1,0.5,0.35 and 0.4 at Al-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively,that means Modern Medical Centrehad scored the highestresolution images, the matter that affects the image quality positivly. Linearity was found to be 74.3%, 83.54%, 90.42% and 88.45% forAl-Taheel Hospital Madani,Ribat University Hospital,Modern Medical Centre,Khartoum Diagnostic Center respectively, that means Modern Medical Centrehad scored the highest linearity measurement the matter that affects the image quality positivly en_US
dc.description.sponsorship Sudan University of Science and Technology en_US
dc.language.iso en en_US
dc.publisher Sudan University of Science and Technology en_US
dc.subject College of Medical en_US
dc.subject Radiologic Science en_US
dc.subject Evaluation of Quality en_US
dc.subject Control of MRI Machines en_US
dc.title Evaluation of Quality Control of MRI Machines. en_US
dc.title.alternative تقويم ضبط الجودة لاجهزة الرنين المغنطيسي en_US
dc.type Thesis en_US