Please use this identifier to cite or link to this item: https://repository.sustech.edu/handle/123456789/25692
Full metadata record
DC FieldValueLanguage
dc.contributor.authorHassan, Omyma Ahmed Eljed-
dc.contributor.authorSupervisor, -Salah Ali Fadlalla-
dc.date.accessioned2021-02-18T12:03:04Z-
dc.date.available2021-02-18T12:03:04Z-
dc.date.issued2017-12-22-
dc.identifier.citationHassan, Omyma Ahmed Eljed .Assessment of Quality Control of X-ray Mammography Machine in Alnilain Medical Center& Royal Care Hospital \ Omyma Ahmed Eljed Hassan ; Salah Ali Fadlalla .- Khartoum:Sudan University of Science and Technology,College of Medical Radiologic Science,2017.- 80p.:ill.;28cm.-M.Scen_US
dc.identifier.urihttp://repository.sustech.edu/handle/123456789/25692-
dc.descriptionThesisen_US
dc.description.abstractA systematic approach for assessing critical performance indicators can be achieved through the implementation of a quality assurance (QA) program. The objective of this study was to evaluate the quality control of x-ray mammography machine in two centers, namely: during the period from October to December 2017. The test results were done by quality control tools and image quality tools. Series of tests(specific radiation output, output variation with kVp, output variation with mAs, half value layer, mean glandular dose, image quality evaluation ) were done using different tools (Semiconductor detector ,AL sheets (0.1 mm thickness) PMMA thickness, image quality tools(TORMAM,TORMAX)). All the results from the two centers were compatible with the international standards values. The study proposed some recommendations which could be useful in this field if applied properly. Future studies in this field should include other QC tests on more clinical centers to guarantee more reliable results.en_US
dc.description.sponsorshipSudan University of Science & Technologyen_US
dc.language.isoenen_US
dc.publisherSudan University of Science and Technologyen_US
dc.subjectMedical Radiologic Sciencesen_US
dc.subjectMedical physicsen_US
dc.subjectQuality Control of X-rayen_US
dc.subjectMammography Machineen_US
dc.subjectAlnilain Medical Centeren_US
dc.titleAssessment of Quality Control of X-ray Mammography Machine in Alnilain Medical Center& Royal Care Hospitalen_US
dc.typeThesisen_US
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Assessment of Quality ....pdfResearch2.38 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.