dc.identifier.citationIbraheem, Rofida Saeed .Quality Control of conventional diagnostic x-ray unit \ Rofida Saeed Ibraheem ; Hussein Ahmed Hassan .- Khartoum:Sudan University of Science and Technology,College of Medical Radiologic Science,2020.- 47p.:ill.;28cm.-M.Scen_US
dc.identifier.urihttp://repository.sustech.edu/handle/123456789/25686-
dc.descriptionThesisen_US
dc.description.abstractThe main objective of this study to assess the performance of x-ray equipments in three radiology department. The study was conducted at radiology department, performed the basic quality control tests Kv accuracy, x ray tube reproduciability, timer accuracy, mAs linearity and half value layer. The results of this study showed that, KVp accuracy, x ray reproducibility, timer accuracy, timer reproducibility, mAs linearity and half value layer were within acceptable limits to two x.ray mashines, thrid mashine fail on most of the QC test. The study recommended that, routine checking and quality testes to x-ray units should be perform in regular and opertaor should be caution when dealing with radiography devices during operation and storage to identify any faults.en_US
dc.description.sponsorshipSudan University of Science & Technologyen_US
dc.language.isoenen_US
dc.publisherSudan University of Science and Technologyen_US
dc.subjectMedical Radiologic Sciencesen_US
dc.subjectMedical physicsen_US
dc.subjectQuality Controlen_US
dc.subjectconventional diagnostic x-ray uniten_US
dc.titleQuality Control of conventional diagnostic x-ray uniten_US
dc.title.alternativeضبط الجود ة لوحدات الاشعة السينية التشخيصية التقليديةen_US
dc.typeThesisen_US
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Quality Control of conventional .....pdfResearch567.71 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.