Please use this identifier to cite or link to this item:
https://repository.sustech.edu/handle/123456789/22698
Title: | Assessment of KVp accuracy of Machines in X-ray Departments of Khartoum state |
Other Titles: | تقييم دقة الكيلو فولتية لاجهزة الاشعة السينية في ولاية الخرطوم |
Authors: | Aldikhari, Nada Mohamed Abdalla Supervisor, -Hussain ahmed Hassan |
Keywords: | Medical Radiologic Sciences Medical Physics KVp accuracy Machines in X-ray |
Issue Date: | 8-Dec-2018 |
Publisher: | Sudan University of Science and Technology |
Citation: | Aldikhari, Nada Mohamed Abdalla . Assessment of KVp accuracy of Machines in X-ray Departments of Khartoum state \ Nada Mohamed Abdalla Aldikhari ; Hussain ahmed Hassan .- Khartoum : Sudan University of Science and Technology , College of Medical Radiologic Science , 2018 .- 63 p. :ill;28cm.- M.Sc |
Abstract: | This study aimed to assessment of KVp accuracy for X-ray departments in Khartoum state . A total of 13 X-ray units in radiology departments in Khartoum state were tested by KVp meter tool . In this study the final resalts were analayzed by Microsoft office Exclle program , The results showed that the acceptable percentage of for X-ray machines was (85%) , greater than unacceptable for X-ray machines (15%) . The final assessment of KVp acceptable accuracy for X-ray departments was most in Khartoum state. It means that implementation of QC program concerning KVp accuracy is very important to the production of image of good quality. |
Description: | Thesis |
URI: | http://repository.sustech.edu/handle/123456789/22698 |
Appears in Collections: | Masters Dissertations : Medical Radiologic Science |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Assessment of KVp ... .pdf Restricted Access | Research | 1.55 MB | Adobe PDF | View/Open Request a copy |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.