Please use this identifier to cite or link to this item: https://repository.sustech.edu/handle/123456789/16633
Full metadata record
DC FieldValueLanguage
dc.contributor.authorAli , Mahmoud Omer Mohamed
dc.contributor.authorHusein , Eltahir Mohmmed
dc.date.accessioned2017-04-25T07:41:20Z
dc.date.available2017-04-25T07:41:20Z
dc.date.issued2015
dc.identifier.citationAli , Mahmoud Omer Mohamed . The Accuracy of Temperature Monitoring of the Incubator for Newborns \ Mahmoud Omer Mohamed Ali , Eltahir Mohmmed Husein .- Journal of Engineering and Computer Sciences (ECS) .- vol 16 , no3.- 2015.- articleen_US
dc.identifier.issnISSN 1605-427X
dc.identifier.urihttp://repository.sustech.edu/handle/123456789/16633
dc.descriptionarticleen_US
dc.description.abstractThe objective of this paper is to investigate the performance of the Incubator for Newborns at the departments of intensive care in Jafer Ibn Aouf Reference Children Hospital, is governmental Children specialized hospital. Using the INCUTM Incubator Analyzer to analyzing various aspects of performance of neonatal incubator and its compliance with the set requirements and regulations according to the technical standard IEC (International Electro technical Commission) 60601-2-19. The data used in this study were obtained by interpreting of the performance test results of neonatal incubator. Numerically, According to the results presented in Tables (1 & 3), the neonatal incubator have achieved acceptable performance and sustaining the requirements of the standard IEC 60601-2-19. In other words, the temperature at points T1, T2, T3, and T4 does not differ from the neonatal incubator average temperature by more than 0.8°C, and this temperature does not differ more than 1.5 °C from the incubator control temperature. Compared to the established technique, the estimation error is negligible.en_US
dc.description.sponsorshipSudan University of Science and Technologyen_US
dc.language.isoen_USen_US
dc.publisherSudan University of Science and Technologyen_US
dc.subjectNeonatal Incubator; Temperature; Performance Test.en_US
dc.titleThe Accuracy of Temperature Monitoring of the Incubator for Newbornsen_US
dc.typeArticleen_US
Appears in Collections:Volume 16 No. 3

Files in This Item:
File SizeFormat 
The Accuracy of Temperature ... .pdf464.05 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.