Please use this identifier to cite or link to this item: https://repository.sustech.edu/handle/123456789/11189
Full metadata record
DC FieldValueLanguage
dc.contributor.authorAli, Ibrahim Galal Alkhlifa
dc.contributor.authorSupervisor,-Hussein Ahmed Hassan
dc.contributor.author
dc.date.accessioned2015-06-25T08:20:52Z
dc.date.available2015-06-25T08:20:52Z
dc.date.issued2015-06-01
dc.identifier.citationAli,Ibrahim Galal Alkhlifa . Assessment of Conventional X-Ray Machine Anode Heel Effect\Ibrahim Galal Alkhlifa Ali ; Hussein Ahmed Hassan.- Khartoum : Sudan University of Science and Technology ,College of Medical Radiologic Science , 2015.-64P. :ill. ;28cm.-M.SC.en_US
dc.identifier.urihttp://repository.sustech.edu/handle/123456789/11189
dc.descriptionThesisen_US
dc.description.abstractThis study used to evaluate the effect of anode heel in conventional X-ray tube Abstrac in image quality femur and humerus dry phantom are exposed, the image was obtained using dry phantom in the Modern Medical Center, Antalia Center , Ear Nose and throat hospital with marks in anode side and cathode side. A total of 10 images was obtained using dry phantom, the exposure factors used to obtain the image, 50 (Kvp), 5 current time product (mAs), and the ESDs was established at a reference point of 100 cm from tube focus for the range of current time product (mAs) and tube potentials encountered in clinical practice. The main results were that the densities increase toward cathode except in digital radiography, which are the same identify in literature, also the variation in densities is due to the variation in x-rays intensities but this does not affect on image quality especially in DR, due to wide capability of post processing of the image which result in enhancing image contrast and visibility and reduced the effect of variation. IV x-ray intensitiesen_US
dc.description.sponsorshipSudan University of Science and Technologyen_US
dc.language.isoenen_US
dc.publisherSudan University of Science and Technologyen_US
dc.subjectX-Rayen_US
dc.subjectConventionalen_US
dc.subjectMachine Anodeen_US
dc.subjectHeelen_US
dc.subjectAssessmenten_US
dc.titleAssessment of Conventional X-Ray Machine Anode Heel Effecten_US
dc.title.alternative‫تقييم تأثير كعب النود على أنبوب الشعة‬ ‫السينية‬en_US
dc.typeThesisen_US
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Assessment of Conventional ....pdfTitle36.77 kBAdobe PDFView/Open
Abstract.pdfAbstract115.53 kBAdobe PDFView/Open
Research.pdfResearch804.45 kBAdobe PDFView/Open