dc.date.issued | 2013-07-01 | |
dc.identifier.citation | Altaib ,Hifa Alhadi Hassan.Assessment of Radiation Dose Output versus Angle Differential in Co-60 Teletherapy:Hifa Alhadi Hassan Altaib ;. Mohammed Ahmed Ali Omer.-khartuom:Sudan University of science & Technology,College of Medical Radiologic Science ,2013.-28p.:ill,:28cm.-M.Sc. | en_US |
dc.identifier.uri | http://repository.sustech.edu/handle/123456789/10306 | |
dc.description | Thesis | en_US |
dc.description.abstract | This study was intended to assess the radiation dose output versus angle differential in Co-60 teletherapy .the radiation output was measured at RICK from cobalt 60 machine by used an ionization chamber and set different SSDs, field sizesand the angleθ, the dose was decreases with angle increase represented by this equations: Y=1.698x+236.5 , Y=-9.337x+294.2 Y=-10.06x+284.3 , Y=-9.996x+260.5 Y=-8.76x+210.8 The data was collected shows the dose increases following the field size increment anddose has been decreases following the SSD increment. The correlation between field size in cm and dose in mGy with angle factor are fitted in the following equation:y= 63.6 − 32.5 , where x refers to field size and y refers to dose in mGy. | en_US |
dc.description.sponsorship | Sudan University of Science and Technology | en_US |
dc.language.iso | en | en_US |
dc.publisher | Sudan University of science & Technology | en_US |
dc.subject | College of Radiology | en_US |
dc.subject | Medical Physics | en_US |
dc.subject | Assessment of Radiation Dose Output versus Angle Differential | en_US |
dc.subject | in Co-60 Teletherapy | en_US |
dc.title | Assessment of Radiation Dose Output versus Angle Differential in Co 60 Teletherapy | en_US |
dc.title.alternative | تقييم المعدل الزاوي لخرج جهاز الكوبالت 60 للعلاج عن بعد | en_US |
dc.type | Thesis | en_US |
Appears in Collections: | Masters Dissertations : Medical Radiologic Science |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Sudan University of ... .pdf Restricted Access | Title | 51.53 kB | Adobe PDF | View/Open Request a copy |
Abstract.pdf Restricted Access | Abstract. | 214.44 kB | Adobe PDF | View/Open Request a copy |
Research .pdf Restricted Access | Research | 1.42 MB | Adobe PDF | View/Open Request a copy |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.