dc.identifier.citationAltaib ,Hifa Alhadi Hassan.Assessment of Radiation Dose Output versus Angle Differential in Co-60 Teletherapy:Hifa Alhadi Hassan Altaib ;. Mohammed Ahmed Ali Omer.-khartuom:Sudan University of science & Technology,College of Medical Radiologic Science ,2013.-28p.:ill,:28cm.-M.Sc.en_US
dc.identifier.urihttp://repository.sustech.edu/handle/123456789/10306
dc.descriptionThesisen_US
dc.description.abstractThis study was intended to assess the radiation dose output versus angle differential in Co-60 teletherapy .the radiation output was measured at RICK from cobalt 60 machine by used an ionization chamber and set different SSDs, field sizesand the angleθ, the dose was decreases with angle increase represented by this equations: Y=1.698x+236.5 , Y=-9.337x+294.2 Y=-10.06x+284.3 , Y=-9.996x+260.5 Y=-8.76x+210.8 The data was collected shows the dose increases following the field size increment anddose has been decreases following the SSD increment. The correlation between field size in cm and dose in mGy with angle factor are fitted in the following equation:y= 63.6􀝔 − 32.5 , where x refers to field size and y refers to dose in mGy.en_US
dc.description.sponsorshipSudan University of Science and Technologyen_US
dc.language.isoenen_US
dc.publisherSudan University of science & Technologyen_US
dc.subjectCollege of Radiologyen_US
dc.subjectMedical Physicsen_US
dc.subjectAssessment of Radiation Dose Output versus Angle Differentialen_US
dc.subjectin Co-60 Teletherapyen_US
dc.titleAssessment of Radiation Dose Output versus Angle Differential in Co 60 Teletherapyen_US
dc.title.alternativeتقييم المعدل الزاوي لخرج جهاز الكوبالت 60 للعلاج عن بعدen_US
dc.typeThesisen_US
Appears in Collections:Masters Dissertations : Medical Radiologic Science

Files in This Item:
File Description SizeFormat 
Sudan University of ... .pdf
  Restricted Access
Title51.53 kBAdobe PDFView/Open Request a copy
Abstract.pdf
  Restricted Access
Abstract.214.44 kBAdobe PDFView/Open Request a copy
Research .pdf
  Restricted Access
Research1.42 MBAdobe PDFView/Open Request a copy


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.